Issue / Malfunction Required Repair
No water to clean lenses; scope is not insufflating. Air/Water Nozzle Cleaning or Replacement; Air/Water Channel Replacement
Control knobs are tight; scope is not angulating properly. Angulation Adjustment or Complete Angulation System Rebuild
Working channel / biopsy channel is blocked. Biopsy Channel Replacement
Scope angulates tightly; insertion tube is snaking. Cable Sheath / Coil Pipe Replacement or Angulation System Rebuild
Scope is filled with fluid; dark liquid leaking from scope; static in image; video buttons not operating. Major Fluid Contamination Repair and Internal Component Restoration
Image is dark; visibility is poor. Fiber Optic Light Guide Replacement
Light guide tube is buckled, leaking, or damaged. Light Guide Tube Replacement
Video buttons #1-5 are not functioning. Video Head Replacement
Image quality issues; static in image; image cuts out when scope is angulated. CCD Chip Repair or Replacement
Insertion tube is buckled, crushed, leaking, or excessively worn. Insertion Tube Replacement